Aligning Sensors to 6 Degrees of Freedom

The tools described in the following sections are only intended to be used with G2 sensors.

Performing alignment using the Surface Align Ring or Surface Align Wide tools (which results in 6 degrees of freedom) involves considerable setup effort. First, the calibration targets are more difficult to manufacture than a calibration bar. Second, the alignment tools have several parameters that must be properly configured to successfully perform an alignment.

When you use the Surface Align Ring or Surface Align Wide tools to create the configuration file containing the sensor transformation, always remember to uncheck Enable uniform spacing in the Scan mode section on the Scan page, even if you intend to use uniform spacing in your application. Also enable Multi-Sensor Array Output.

Aligning a system of sensors to 6 degrees of freedom involves the use of one of two Surface measurement tools (Surface Align Wide or Surface Align Ring). These tools create a set of transformations and store them in a configuration file. The resulting alignment is more accurate compared to the other methods available on the Alignment panel, and includes compensations for X angle rotations. Note that in order to apply the transformations to scan data, you must use a "stitching" tool that corresponds to the tool used to create the transformations.

Both tools produce configuration files, and can optionally load previously saved configuration files, which is useful if you need to set the system up again. These configuration files are stored in C:\GoTools\SurfaceAlign\.